ORIGINAL ARTICLE
Year : 2012  |  Volume : 27  |  Issue : 2  |  Page : 89-94  

Spreadsheet program for estimating recovery coefficient to get partial volume corrected standardized uptake value in clinical positron emission tomography-computed tomography studies


Department of Nuclear Medicine, All India Institute of Medical Sciences, New Delhi, India

Date of Web Publication18-Apr-2013

Correspondence Address:
Rakesh Kumar
Department of Nuclear Medicine, All India Institute of Medical Sciences, New Delhi - 110 029
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0972-3919.110688

Rights and Permissions
   Abstract 

Purpose: To develop a spreadsheet program for estimation of recovery coefficient (RC) to get partial volume corrected (PVC) standardized uptake value (SUV) in clinical positron emission tomography-computed tomography (PET-CT) studies. Materials and Methods: For formulation of this program we used data from a phantom study conducted at our center in which a phantom with a sphere assembly (seven spheres-different diameters) was filled with 18F-Fluorodeoxyglucose solution to get a sphere/background ratio of 8:1, 10:1 and 12:1. PET-CT images were acquired. RC was then calculated from processed PET-CT images. We plotted graph of RC versus lesion-size at different sphere/background ratio using MS Excel function. There was logarithmic increase in RC with increase in lesion size. We fitted the data with a logarithmic equation and found optimum fit (least-square fit).We then validated this program with clinical data using 42 lung nodules in five patients. Results : The program estimates the value of RC and object to background ratio in PET-CT for the input lesion-size and displays graph with trendline. When the user enters SUV and background activity measured in clinical PET-CT, it provides the value of RC and PVC SUV. It also validates the data entry and displays appropriate message. It is consistent, reproducible, accurate and provides output for wide range of lesion-sizes (71% of lesions evaluated); however, program does not give output for lesion-size < 9 mm. Conclusion: The present spreadsheet program is a useful and easy tool for calculating PVC SUV of clinical PET-CT lesions.

Keywords: Partial volume correction, positron emission tomography-computed tomography, spreadsheet program, standardized uptake value


How to cite this article:
Pandey AK, Sharma P, Pandey M, Aswathi K K, Malhotra A, Kumar R. Spreadsheet program for estimating recovery coefficient to get partial volume corrected standardized uptake value in clinical positron emission tomography-computed tomography studies. Indian J Nucl Med 2012;27:89-94

How to cite this URL:
Pandey AK, Sharma P, Pandey M, Aswathi K K, Malhotra A, Kumar R. Spreadsheet program for estimating recovery coefficient to get partial volume corrected standardized uptake value in clinical positron emission tomography-computed tomography studies. Indian J Nucl Med [serial online] 2012 [cited 2022 Jan 20];27:89-94. Available from: https://www.ijnm.in/text.asp?2012/27/2/89/110688


   Introduction Top


Standardized uptake value (SUV) is the most widely used semi-quantitative parameter used in positron emission tomography-computed tomography (PET-CT). It is used for categorizing malignant versus benign lesions, and a max SUV of 2.5 or greater is usually consistent with malignancy. [1],[2] SUV is also being employed for response monitoring and prognostication of wide array of tumors. [3],[4],[5],[6],[7],[8],[9],[10],[11],[12]

However, assessment of SUV is erroneous in small lesions. For such lesions, if the characterization of a lesion is based on a SUV cut-off (suppose SUV max > 2.5), the results can be radically altered. The lesions smaller than twice or thrice of full width half maximum (FWHM) of PET scanner cannot display its true structural (true size) and functional properties (true counts) because of "partial volume effect" (PVE). [13] Therefore, small-sized lesions on PET-CT should be corrected for the error introduced by PVE. This correction can be achieved by applying a particular factor called the recovery coefficient (RC). [14],[15],[16],[17],[18],[19]

We had previously carried out a phantom study at our institute to evaluate the PVE introduced in spheres of a phantom and the RC values to be applied for partial volume corrected (PVC) of those spheres. [20] Based on a that study, we had generated a "Look Up Table" for seven different size of lesion only, namely 11, 13, 13.5, 14.4, 15, 18.3 and 19.3 mm having object to background ratio in the PET scan ranging from 2.70 to 19.60, as was seen in the phantom. In routine clinical practice, however, we can encounter lesions of variable sizes and a wide range of lesion to background ratios. Hence, there was a need for a program that can provide the value of RC for all lesion sizes having all possible value of object to background ratios on routine clinical PET-CT. We have developed such a spreadsheet program, based on our experimental data.


   Materials and Methods Top


Phantom study

The phantom images were acquired using a dedicated PET-CT scanner (Biograph 2, Siemens Medical Solutions, Erlangen, Germany). The phantom was fabricated locally using tissue equivalent material. It consisted of two parts: A hollow cylinder having diameter 18.5 cm and length 20 cm, and multiple spheres. The sphere assembly contained seven spheres of varying diameter ranging from 11 mm to 19.3 mm (i.e., 11, 13, 13.5, 14.4, 15, 18.3 and 19.3 mm). The phantom was filled with water containing 18F-Fluorodeoxyglucose (FDG) to make the sphere versus background activity concentration (μC i /ml) ratio 8:1. A circular region of interest (ROI) of size as estimated by CT was drawn on PET images of hot spheres. The maximum activity concentration (μC i /ml) for the hot sphere was noted in all the slices where the image of that sphere was seen on PET-CT. Slice containing maximum activity concentration was determined for each sphere. For measuring the background activity concentration, 12 different circular regions with same area were defined. The processes of drawing region of interest ROIs were repeated on two transverse slices above and two transverse slices below the maximum intensity pixel. Mean activity concentration was recorded for each region and then the average for all 60 regions was calculated.

The value of RC for each sphere size was calculated by the formula:



The same experiment was repeated at sphere/background activity ratios of 10:1 and 12:1.

Spreadsheet program

This program was created in Microsoft Excel spreadsheet program (Microsoft Office, Microsoft Corp., New Jersey, USA) using the personal computer having Pentium (R) 4 CPU 3.00 GHz, 2.99 GHz, and 504 MB of RAM. We plotted graph of RC versus lesion-size (diameter of sphere in mm) with sphere to background activity ratio at the time of acquisition (namely, 8:1, 10:1 and 12:1). As size of lesion increased, there was an increase in the value of RC and this increase was logarithmic. A logarithmic data fit was performed which provided the following three equations involving RC and lesion-size (diameter of sphere):



Where y = RC and x = lesion-size (diameter of sphere in mm).

Similarly, we again plotted three separate graphs of measured object to background ratio in the PET scan (max pixel in hot sphere: Average background pixel) versus lesion-size (diameter of sphere), and logarithmic data fit were performed. The following three different equations were available for object to background ratios of 8:1, 10:1, and 12:1:



Where y = measured object to background ratio on PET and x = lesion-size (diameter of sphere in mm).

We used these three equations (Equations 4, 5 and 6) to estimate the value of measured object to background ratio in the PET scan for a given lesion size when spheres would have been filled with 8:1, 10:1 and 12:1 sphere/background activity ratio at the time of acquisition.

Using equations 1-6, we made the program to generate on-line the value of RC and estimated value of object to background ratio on PET images for a given lesion size when spheres would have been filled with 8:1, 10:1 and 12:1 sphere/background activity ratio at the time of acquisition. The program also generates an on-line a graph, performs logarithmic fit and displays a logarithmic trendline equation for the generated data points.

The program requires the maximum SUV value of the lesion and background as input data from the user. As soon as user enters this data, the program calculates the value of the RC to be applied for the entered lesion-size, based on the logarithmic trendline equation between RC and estimated value of object to background ratio in the PET scan and calculates the corrected SUV using Equation 7:



Clinical validation

Test cases were generated to test this program so that boundary condition for it can be defined in which this program can successfully estimate the value of RC and corrected SUV after partial volume correction. Data of five patients with 42 pulmonary nodules was used for this purpose. The SUV max of these nodules and the background was measured on PET. The size was measured on CT. The RC and the PVC SUV of these nodules was calculated using the spreadsheet program.


   Results Top


Graphical User Interface of the program is shown in [Figure 1]. Data entry has to be completed in two steps. In the first step, the program requires the user to input the lesion-size in mm. Based on the data entered, the program estimates the value of RC and object to background ratio in the PET scan, generates a graph, displays equation for the trendline [see in the upper right end of [Figure 1]]. In the 2 nd step, user is also required to enter the value of SUV of the lesion and background for which the PVE corrected SUV is desired. As soon as the user completes the entry of data, the program displays both the RC to applied and corrected SUV value on result areas [lower right hand side of the [Figure 1]].The program gives consistent and reliable result. It is stable and reproducible. It provides appropriate and clear message when the calculation of RC is not possible with this program. In order to validate this program, data of all 42 lesions-their sizes and their SUV to background ratio were fed as input parameter to the program in order to calculate the RC and PVC SUV value. Result has been given in [Table 1]; [Figure 2]. The lesion-size varied from 5.03 mm to 27.67 mm and SUV to background ratio varied from 1.05 to 19.76. For some lesion-sizes, the generated array of RC and estimated object-to-background ratio was such that it was not possible to calculate the trendline equation (i. e. not possible to fit the data). For the selected 42 lesion-sizes for validation, by using the previous "look up table" generated from phantom study, we were able to estimate the value of RC and PVC SUV for 10 lesion only (24% lesions). However, with the present spreadsheet program we estimated RC and PVC SUV value for 30 lesions (71% lesions).
Figure 1: Graphical user interface of the spreadsheet program

Click here to view
Figure 2: Bar chart displaying the lesion-size (mm) and their standardized uptake value to background ratio of the 42 pulmonary lesions selected for the validation of the spreadsheet program

Click here to view
Table 1: Finding of positron emission tomography-computed tomography of the 42 pulmonary nodules used for validation of the spreadsheet program

Click here to view


It can be seen from the [Table 1] that this program could not estimate the value of RC and PVC SUV for the lesion-size less than 9 mm (total 10 cases out of 42). It has estimated the value of RC more than 1 for the two lesions of lesion-sizes 16.57 mm and 27.66 mm, which are greater than 16 mm. However, at a same lesion-size of 16.57 mm we have also obtained negative value of RC (-0.263) and corrected SUV value (-2.91). For some lesion-sizes, the generated array of RC and estimated object-to-background ratio was such that it was not possible to calculate the trendline equation (it was not possible to fit the data). In such cases, the program displays the error message "some trendlines cannot be calculated from data containing negative or zero value" [Figure 3].

For lesion-size less than 18 mm, there is a PVE. Based on data fit, however in some cases, the estimated RC value was negative. If we use this value for PVC SUV value then it will be less than measured SUV value without any correction applied. This is erroneous and not acceptable. Hence, user must reject such values.
Figure 3: Error message generated in the spreadsheet program when it is not possible to estimate recovery coefficient and partial volume corrected standardized uptake value

Click here to view



   Discussion Top


There was a need for a program that can provide the value of RC for all lesion sizes having all possible value of object to background ratio in PET-CT, encountered in the routine clinical practice. We developed such a spreadsheet program based on our experimental data. The program requires Microsoft Excel 2003 to execute. The program is robust and very useful, time saving, and eliminates manual calculation error, which might be introduced in the process of calculation of PVC SUV.

Recently, some studies [21],[22] in the literature have generated "look up table" for RC and partial volume correction based upon phantom study. These were similar to our phantom study but with difference in the lesion-size and object-to-background ratio. Furthermore, our study was hot sphere in hot background model while most of the studies employed hot sphere in cold background model. By using the "look up tables" obtained from the phantom studies only those clinical lesions whose size matches the value of the lesion size and the measured object-to-background ratio with that of the study can be corrected for PVE. Our spreadsheet program is entirely different from a simple "look up table"; here we were finding the RC for a given range of lesion-size (11-19.3 mm) in the various condition of object-to-background ratio including the object-to-background ratio not available in simple look table.

However, this program has certain limitations. As RC depends on various factors that can affect the RC and influence the magnitude of PVEs in most cases, therefore, this program can give accurate results only if certain conditions are met. Our program could not estimate the value of RC and PVC SUV for the lesion-size less than 9 mm (total 10 cases out of 42), because the generated data was so dispersed that it was not possible to fit the data for logarithmic relationship between RC and estimated value of object to background ratio. It has estimated the value of RC more than 1 for the two lesions of lesion-sizes (16.57 mm and 27.66 mm, which are greater than 16 mm). However, at a same lesion-size of 16.57 mm we have also obtained negative value of RC (-0.263) and corrected SUV value (-2.91). Serial No. 18 has SUV/Background ratio 19.76 while Serial No. 42 has SUV/Background ratio 2.02, and both have same lesion-size as 16.57. This difference is obviously due to SUV/Background ratio and based on our experimental data it is correct because for same lesion size we had the less value as object-to-background ratio in the experimental data increases that is from 8:1 to 12:1, the value of RC and measured value of object-to-background ratio in PET scan decreases. Therefore, for 19.7 SUV to background ratio the value of RC will be more in comparison to SUV to background ratio of 2.02.

Since this spreadsheet program is based on the data collected in our previous study, [20] any user, who has similar experimental conditions, can use this program. In brief, the following conditions need to be satisfied:

  • The FWHM of the PET system should be 6 mm.
  • Geometry of lesion should be spherical.
  • The maximum pixel value within a ROI should be taken to calculate the value of RCs.

   Conclusion Top


We have presented a spreadsheet program which can provide PVC SUV of clinical lesions with ease and accuracy. This can be used for most but not all lesions seen in clinical PET-CT imaging.

 
   References Top

1.Coleman RE, Laymon CM, Turkington TG. FDG imaging of lung nodules: A phantom study comparing SPECT, camera-based PET, and dedicated PET. Radiology 1999;210:823-8.  Back to cited text no. 1
    
2.Shreve PD, Steventon RS, Deters EC, Kison PV, Gross MD, Wahl RL. Oncologic diagnosis with 2-fluorine-18 fluoro-2-deoxy-D-glucose imaging: Dual-head coincidence gamma camera versus positron emission tomographic scanner. Radiology 1998;207:431-7.  Back to cited text no. 2
    
3.Adler LP, Blair HF, Makley JT, Williams RP, Joyce MJ, Leisure G, et al. Noninvasive grading of musculoskeletal tumors using PET. J Nucl Med 1991;32:1508-12.  Back to cited text no. 3
    
4.Buonocore E. Comparison of PET with conventional imaging techniques. In: Hubner KF, Collmann J, Buonocore F, Kabalka GW, editors. Clinical Positron Emission Tomography. St. Louis. MO: Mosby Year Book; 1992;17-22.  Back to cited text no. 4
    
5.Adler LP, Crowe JP, al-Kaisi NK, Sunshine JL. Evaluation of breast masses and axillary lymph nodes with F-18 2-deoxy-2-fluoro-D-glucose PET. Radiology 1993;187:743-50.  Back to cited text no. 5
    
6.Laubenbacher C, Saumweber D, Wagner-Manslau C, Kau RJ, Herz M, Avril N, et al. Comparison of fluorine-18-fluorodeoxyglucose PET, MRI and endoscopy for staging head and neck squamous-cell carcinomas. J Nucl Med 1995;36:1747-57.  Back to cited text no. 6
    
7.Rodriguez M, Rehn S, Ahlström H, Sundström C, Glimelius B. Predicting malignancy grade with PET in non-Hodgkin's lymphoma. J Nucl Med 1995;36:1790-6.  Back to cited text no. 7
    
8.Dehdashti F, Mortimer JE, Siegel BA, Griffeth LK, Bonasera TJ, Fusselman MJ, et al. Positron tomographic assessment of estrogen receptors in breast cancer: Comparison with FDG-PET and in vitro receptor assays. J Nucl Med 1995;36:1766-74.  Back to cited text no. 8
    
9.Avril N, Dose J, Jänicke F, Ziegler S, Römer W, Weber W, et al. Assessment of axillary lymph node involvement in breast cancer patients with positron emission tomography using radiolabeled 2-(fluorine-18)-fluoro-2-deoxy-D-glucose. J Natl Cancer Inst 1996;88:1204-9.  Back to cited text no. 9
    
10.Vitola JV, Delbeke D, Meranze SG, Mazer MJ, Pinson CW. Positron emission tomography with F-18-fluorodeoxyglucose to evaluate the results of hepatic chemoembolization. Cancer 1996;78:2216-22.  Back to cited text no. 10
    
11.Rasey JS, Koh WJ, Evans ML, Peterson LM, Lewellen TK, Graham MM, et al. Quantifying regional hypoxia in human tumors with positron emission tomography of 18F fluoromisonidazole: A pretherapy study of 37 patients. Int J Radiat Oncol Biol Phys 1996;36:417-28.  Back to cited text no. 11
    
12.Hübner KF, Buonocore E, Gould HR, Thie J, Smith GT, Stephens S, et al. Differentiating benign from malignant lung lesions using "quantitative" parameters of FDG PET images. Clin Nucl Med 1996;21:941-9.  Back to cited text no. 12
    
13.Gambhir S. Quantitation of the physical factors affecting the tracer kinetic modeling cardiac positron emission tomography data [dissertation]. Los Angeles, CA: University of California at Los Angeles; 1990.  Back to cited text no. 13
    
14.Zasadny KR, Wahl RL. Standardized uptake values of normal tissues at PET with 2-fluorine-18-fluoro-2-deoxy-D-glucose: Variations with body weight and a method for correction. Radiology 1993;189:847-50.  Back to cited text no. 14
    
15.Kim CK, Gupta NC, Chandramouli B, Alavi A. Standardized uptake values of FDG: Body surface area correction is preferable to body weight correction. J Nucl Med 1994;35:164-7.  Back to cited text no. 15
    
16.Kessler RM, Ellis JR Jr, Eden M. Analysis of emission tomographic scan data: Limitations imposed by resolution and background. J Comput Assist Tomogr 1984;8:514-22.  Back to cited text no. 16
    
17.Hoffman EJ, Huang SC, Phelps ME. Quantitation in positron emission computed tomography: 1. Effect of object size. J Comput Assist tomography 1979;3:299-308.  Back to cited text no. 17
    
18.Mazziotta JC, Phelps ME, Plummer D, Kuhl DE. Quantitation in positron emission computed tomography: 5. Physical - Anatomical effects. J Comput Assist Tomogr 1981;5:734-43.  Back to cited text no. 18
    
19.Chen CH, Muzic RF Jr, Nelson AD, Adler LP. Simultaneous recovery of size and radioactivity concentration of small spheroids with PET data. J Nucl Med 1999;40:118-30.  Back to cited text no. 19
    
20.Pandey AK, Pandey M, Sharma P, Aswathi KK, Gupta A, Malhotra A, et al. A phantom study for evaluation of partial volume effect and estimation of recovery coefficient for small sized lesion on PET-CT. Ann Nucl Med 2012;[In press].  Back to cited text no. 20
    
21.Srinivas SM, Dhurairaj T, Basu S, Bural G, Surti S, Alavi A. A recovery coefficient method for partial volume correction of PET images. Ann Nucl Med 2009;23:341-8.  Back to cited text no. 21
    
22.Miwa K, Hayasaka K, Umeda T, Kojima H. Partial volume corrected SUV for monitoring the response of non-Hodgkin lymphoma on 18FDG-PET/CT. J Nucl Med 2010;51:S2072.  Back to cited text no. 22
    


    Figures

  [Figure 1], [Figure 2], [Figure 3]
 
 
    Tables

  [Table 1]


This article has been cited by
1 Does the beta regularization parameter of bayesian penalized likelihood reconstruction always affect the quantification accuracy and image quality of positron emission tomography computed tomography?
Zhifang Wu, Binwei Guo, Bin Huang, Bin Zhao, Zhixing Qin, Xinzhong Hao, Meng Liang, Jun Xie, Sijin Li
Journal of Applied Clinical Medical Physics. 2021; 22(3): 224
[Pubmed] | [DOI]
2 Influence of reconstruction parameters of positron emission tomograph scanning on the effect of partial volume of the pathological lesion
E. V. Emelyanenko, I. G. Tarutin, P. A. Belobokov
Proceedings of the National Academy of Sciences of Belarus, Physical-Technical Series. 2021; 66(3): 356
[Pubmed] | [DOI]
3 Methodology and hardware for assessing the quantitative characteristics of pet images in the study of dynamic objects
E. V. Emelyanenko, M. N. Piatkevich, I. G. Tarutin
Proceedings of the National Academy of Sciences of Belarus, Physical-Technical Series. 2021; 66(4): 496
[Pubmed] | [DOI]



 

Top
  
 
  Search
 
  
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
    Access Statistics
    Email Alert *
    Add to My List *
* Registration required (free)  

 
  In this article
    Abstract
   Introduction
    Materials and Me...
   Results
   Discussion
   Conclusion
    References
    Article Figures
    Article Tables

 Article Access Statistics
    Viewed1538    
    Printed68    
    Emailed1    
    PDF Downloaded149    
    Comments [Add]    
    Cited by others 3    

Recommend this journal