Year : 2022  |  Volume : 37  |  Issue : 2  |  Page : 113-120

Evaluation of reconstruction algorithms to validate the nema phantom results in clinical scenario – A comparative study using time-of-flight versus non-time-of-flight positron emission tomography imaging


Department of Nuclear Medicine, Post Graduate Institute of Medical Education and Research, Chandigarh, India

Correspondence Address:
Dr. Baljinder Singh
Department of Nuclear Medicine, Post Graduate Institute of Medical Education and Research, Chandigarh - 160 012
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/ijnm.ijnm_137_21

Rights and Permissions

Objectives: The objective is to standardize the reconstruction parameters for the time-of-flight (TOF) versus non-TOF positron-emission tomography/computed tomography (PET/CT) imaging data and validation of the same in a clinical setting. Methods: The four spheres (10.0/13.0/17.0/22.0 mm) of the PET phantom (NEMA IQ Nu 2-2001) were filled with four times higher activity of [18F]-NaF than the background (5.3kBq/mL). Imaging (image matrix – 128 × 128 × 47, 2 min, 3D model) was done using two different (TOF/non-TOF) PET scanners. Phantom data were reconstructed in TOF and non-TOF modes for lutetium–yttrium oxyorthosilicate and non-TOF mode for bismuth germanate-based PET scanners. The reconstructed data (by varying iteration and subsets) that provided the best image contrast and signal-to-noise ratio (SNR) were evaluated. The whole-body [18F]-fludeoxyglucose (FDG) PET/CT scans (7–8 frames; 2.0 min/frame) in 16 lymphoma patients were acquired at 60 min after injecting the radioactivity (370.0–444.0 MBq of [18F]-FDG. The clinical PET/CT data were reconstructed using phantom-derived reconstruction parameters and evaluated for image contrast and SNR of the detected lesions. Results: TOF reconstruction at second iteration provided significantly (P ≤ 0.02) higher SNR (20.7) and contrast (contrast recovery coefficient/background variability = 3.21) for the smallest hot lesions (10.0 mm) in the phantom than the non-TOF system. Similarly, in patient data analysis for the selected FDG avid lesions, the SNR values were significantly (P = 0.02) higher (13.3 ± 6.49) in TOF than (11 ± 6.48) in non-TOF system. Further, the small (≤10.0 mm) lesions were seen more distinctly in TOF system. Conclusion: It is thus observed that TOF reconstruction converged faster than the non-TOF, and the applicability of the same may impact the image quality and interpretation in the clinical PET data. The validation of the phantom-based experimental reconstruction parameters to clinical PET imaging data is highly warranted.


[FULL TEXT] [PDF]*
Print this article     Email this article
 Next article
 Previous article
 Table of Contents

 Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
 Citation Manager
 Access Statistics
 Reader Comments
 Email Alert *
 Add to My List *
 * Requires registration (Free)
 

 Article Access Statistics
    Viewed460    
    Printed8    
    Emailed0    
    PDF Downloaded77    
    Comments [Add]    

Recommend this journal