Design and evaluation of MeVislab networks for co-registration and cropping of positron emission tomography/computed tomography scans


Anil Kumar Pandey1, Akshima Sharma2, Sunil Kumar1, Ashish Negi1, Satnam Singh1, Sanjay Kumar2, Rakesh Kumar1 
1 Department of Nuclear Medicine, All India Institute of Medical Sciences, New Delhi, India
2 Department of Urology, All India Institute of Medical Sciences, New Delhi, India

Correspondence Address:
Dr. Anil Kumar Pandey
Ph.D, Additional Professor (Medical Physics), Department of Nuclear Medicine, All India Institute of Medical Sciences, Ansari Nagar, New Delhi - 110 029
India

Objective: The aim of the present study was to design and evaluate two MeVisLab networks, one for co-registration of positron emission tomography/computed tomography (PET/CT) images and second for cropping the co-registered PET/CT images. Materials and Methods: Two MeVisLab networks, one to co-register and export PET/CT DICOM images and second for cropping the co-registered PET/CT images were designed using different modules of registration toolkit MERIT. One hundred and twenty-five PET/CT studies were exported from Siemens and GE scanners in DICOM format. These images were co-registered and cropped with our designed networks. The images co-registered with our network were compared visually with the co-registered images of same PET/CT studies on vendor provided workstations by an experienced nuclear medicine physician (NMP). The perfection of the cropping of co-registered images was also assessed visually. Results: Visually, NMP found all 125 images co-registered using the network designed in our study similar to the co-registered images of vendor provided workstations. Furthermore, the cropping of all co-registered images was perfectly done by our network. Conclusion: Two MeVisLab networks designed and evaluated in the present study can be used for co-registration of PET/CT DICOM images and cropping the co-registered PET/CT images.


How to cite this article:
Pandey AK, Sharma A, Kumar S, Negi A, Singh S, Kumar S, Kumar R. Design and evaluation of MeVislab networks for co-registration and cropping of positron emission tomography/computed tomography scans.Indian J Nucl Med 2022;37:37-42


How to cite this URL:
Pandey AK, Sharma A, Kumar S, Negi A, Singh S, Kumar S, Kumar R. Design and evaluation of MeVislab networks for co-registration and cropping of positron emission tomography/computed tomography scans. Indian J Nucl Med [serial online] 2022 [cited 2022 Jul 2 ];37:37-42
Available from: https://www.ijnm.in/article.asp?issn=0972-3919;year=2022;volume=37;issue=1;spage=37;epage=42;aulast=Pandey;type=0