Figure 3: (a) The time activity curve from the radioactivity sensor (γ1) showing peak activity between 10 and 15 min depicting transfer of the activity of Ga-68 eluent to the reactor, (b) time activity curve from the radioactivity sensor (γ2) showing peak activity between 26 to 32 min depicting trapping of the final labelled product on the light C-18 cartridge, (c) time activity graph from the radioactivity sensor (γ3) showing peak activity between 26 to 32 min depicting transfer of the final labelled product to the product vial, (d) temperature-time graph showing peak temperature of 125°C from 14 to 25 min of the radiolabeling process

Figure 3: (a) The time activity curve from the radioactivity sensor (γ1) showing peak activity between 10 and 15 min depicting transfer of the activity of Ga-68 eluent to the reactor, (b) time activity curve from the radioactivity sensor (γ2) showing peak activity between 26 to 32 min depicting trapping of the final labelled product on the light C-18 cartridge, (c) time activity graph from the radioactivity sensor (γ3) showing peak activity between 26 to 32 min depicting transfer of the final labelled product to the product vial, (d) temperature-time graph showing peak temperature of 125°C from 14 to 25 min of the radiolabeling process